Lab Hood and BSC Monitors

For safety and regulatory compliance reasons, it is essential to monitor air flow in laboratory fume hoods, bio-safety cabinets (BSCs), or other types of hoods in critical environments. Alnor AirGard® Lab Hood Monitors and Air Flow Monitors provide visual and audible indications of safe levels of air flow when installed on lab hoods or BSCs.


Designed to meet intent of:

ANSI/AIHA Z9.5-2003
3.3.3 Flow-Measuring Device for Laboratory Chemical Hoods

All hoods shall be equipped with a flow indicator, flow alarm, or face velocity alarm indicator to alert users to improper exhaust flow.
 
NFPA 45-2000
6.8.7 Measuring Device for Hood Airflow
6.8.7.1
- A measuring device for hood airflow shall be provided on each laboratory hood.
6.8.7.2 - The measuring device for hood airflow shall be a permanently installed device and shall provide constant indication to the hood user of adequate or inadequate hood airflow.
 
SEFA 1.2-2002
4.1.10 - All hoods shall have some type of monitor for indicating face velocity or exhaust flow verification… A ribbon taped to the bottom of the sash is not acceptable.
 
CFR Part 1910.1450
Appendix A - National Research Council Recommendations Concerning Chemical Hygiene in Laboratories
C. The Laboratory Facility
4.  Ventilation
    (b) Hoods
…each hood should have a continuous monitoring device to allow convenient confirmation of adequate hood performance before use.
 
NSF-49-2002
5.23 Alarms
5.23.3 Type B exhaust alarm... Audible and visual alarms shall be required to indicate a 20% loss of exhaust volume within 30 seconds...
 


Products

For more information, choose from the following lab hood and air flow monitors:

AirGard® Lab Hood Monitors

Monitors with visual and audible alarms ensure safe levels of air flow in laboratory fume hoods. Excellent for new installations or retrofits.
 
AirGard® Air Flow Monitors

Indicate safe air flows for bio-safety cabinets (BSCs) and other critical environments, including those requiring an explosion proof monitor.
 




Selection Guide

 
Application Lab Hood Face Velocity Lab Hood Face Velocity Lab Hood Face Velocity Lab Hood Face Velocity Air Velocity(Laminar Flow, BioSafety Cabinet) Air Velocity(Laminar Flow, BioSafety Cabinet)
Measurement Range 50-250 ft/minute 50-250 ft/minute 50-250 ft/minute 0.07 – 0.15 in H2O 50-2000 ft/minute 50-2000 ft/minute
Measurement Accuracy 5% of reading 5% of reading 5% of reading ± 0.04 in H2O at min
± 0.06 in H2O at max
5% of reading 5% of reading
Mounting Flush Surface Flush Surface Flush Flush
Alarms Low Low Low, High Low Low Low, High
Visual Indication Normal, Alarm Normal, Alarm Normal, Caution, Alarm Normal, Alarm Normal, Alarm Normal, Caution, Alarm
Audible Alarm Yes Yes Yes Yes Yes Yes
Velocity Display No No Yes No No Yes
Bar-Graph Trend No No Yes No No Yes
Intrinsically Safe No No No Yes No No
Measurement Type Integral Thermal Sensor Integral Thermal Sensor Integral Thermal Sensor Pressure Switch Remote Thermal Sensor Remote Thermal Sensor
Power 24VDC (Plug-In Wall Transformer Included in US) 24VDC (Plug-In Wall Transformer Included in US) 24VDC (Plug-In Wall Transformer Included in US) 120VAC 24VDC (Plug-In Wall Transformer Included in US) 24VDC (Plug-In Wall Transformer Included in US)

 