AirGard® Lab Hood Monitors
Alnor AirGard Lab Hood Monitors indicate safe levels of air flow in laboratory
fume hoods. The Model 200, 335, and 405 Monitors are used to measure and
report on exhaust flow in many critical or controlled indoor environments,
including pharmaceutical and university laboratories. The AirGard 200
and AirGard 405 Monitors are simple, low-cost options for your monitoring
requirements. The AirGard 335 provides the same reliable measurements
with added features that make it more versatile. All these models are
easy to install, making them an excellent choice for both new installations
and retrofits.
The AirGard 200, 335 and 405 Lab Hood Monitors continuously monitor lab
hood face velocity by measuring the air velocity of clean air from the
room passing through the air inlet on the front of the monitor. Maintaining
an adequate face velocity is critical for containment of hazardous substances
that may be present in the lab hood.
When the face velocity drops below a user-determined level, all AirGard
Lab Hood Monitors activate an audible alarm, visual alarm, and relay output.
The 200 and 405 monitors indicate low face velocity with a red LED and
an audible alarm. The 335 monitor has jumbo LEDs to indicate safe, marginal,
or low duct exhaust and sounds an audible alarm during low face velocity
conditions. The AirGard 335 Monitor also has a digital bar graph to indicate
face velocity conditions, an optional digital velocity display to aid
in the diagnosis of problem face velocities, and a high flow alarm to
help detect excessive air velocity. In addition, all AirGard Lab Hood
Monitors can report unsafe face velocity conditions to other personnel
in a remote location by means of a relay output.
Installation and set-up of the Alnor AirGard Lab Hood Monitors is simple
and quick. The 200 and 335 Monitors are flush mounted. The 405 Monitor
is a surface-mounted version ideal for retrofitting existing lab hoods.
Calibration of the AirGard 200 and 405 Monitors consists of setting the
low alarm set point. The AirGard 335 Monitor requires a 2-point velocity
calibration.

| |
200/405
|
335
|
| range |
70-250 fpm (0.35-1.27 m/s) |
50-250 fpm (0.25-1.27 m/s) |
| repeatability |
±10% of reading or 10 fpm (0.05 m/s), whichever
is greater |
±10% of reading or 10 fpm (0.05 m/s), whichever
is greater |
| visual
display |
jumbo green and red LEDs |
bar graph with green, yellow, and
red zones; jumbo green, yellow, and red LEDs |
| alarm indications |
jumbo red LED, audible alarm, relay output |
jumbo red LED, audible alarm, relay output |
| input power |
9 V AC |
9-30 V AC/DC |
| mounting |
flush (200) or surface (405) |
flush |
| agency listings |
CE |
CE, UL |
More detailed specifications
|